FluoroMax
®
-4 & FluoroMax
®
-4P with USB v. B (23 Oct 2009) 
Troubleshooting 
7-9 
Further assistance... 
Read all software and accessory manuals before contacting the Fluorescence Service 
Department. Often the manuals show the problem’s cause and a method of solution. 
Technical support is available for both hardware and software troubleshooting. 
Before contacting the service department, however, complete the following steps. 
If this is the first time the problem has occurred, 
try turning off the system and accessories. 
After a cool-down period, turn everything back on. 
Make sure all accessories are properly confi-
gured, and turned on as needed. 
Following the instructions in Chapter 3, System 
Operation, run a lamp scan and a water Raman 
scan to make sure the system is properly cali-
brated. 
Print the spectrum for each and note the peak intensities. 
Check this chapter to see if the problem is dis-
cussed. 
Try to duplicate the problem and write down the 
steps required to do so. 
The service engineers will try to do the same with a test system. Depending on 
the problem, a service visit may not be required. 
If an error dialog box appears in FluorEs-
sence™, write down the exact error displayed. 
In FluorEssence™, in the FluorEssence main 
window’s toolbar, choose Help: 
A drop-down menu appears. 
Under Help, choose About FluorEssence.... 
This opens the About FluorEssence window: 
Metadata in pdf documents - add, remove, update PDF metadata in C#.net, ASP.NET, MVC, Ajax, WinForms, WPF
Allow C# Developers to Read, Add, Edit, Update and Delete PDF Metadata
clean pdf metadata; get pdf metadata
Metadata in pdf documents - VB.NET PDF metadata library: add, remove, update PDF metadata in vb.net, ASP.NET, MVC, Ajax, WinForms, WPF
Enable VB.NET Users to Read, Write, Edit, Delete and Update PDF Document Metadata
add metadata to pdf; modify pdf metadata
FluoroMax
®
-4 & FluoroMax
®
-4P with USB v. B (23 Oct 2009) 
Troubleshooting 
7-10 
The version of the software (both 
FluorEssence™ and Origin
®
) is 
listed here. 
Click the View Sys-
tem Info button. 
The Installed Components window 
appears, displaying all the software 
required for FluorEssence™.  
10 
Record the informa-
tion by clicking the: 
•  Save To File... button, which 
saves the information to a file; 
•  Zip Info button, which com-
presses the information while sav-
ing it; 
•  Print Info button, which prints 
out the software information. 
11 
Click the OK button 
to close the Installed Components window. 
12 
Click the OK button to close the About FluorEs-
sence window. 
13 
Write down the software’s version numbers, 
along with the purchase dates, model numbers, 
system configuration, and serial numbers of the 
instrument and its accessories. 
14 
Determine the SpectrAcq firmware version: 
C# PDF Library SDK to view, edit, convert, process PDF file for C#
NET empowers C# developers to implement fast and high quality PDF conversions to or from multiple supported images and documents. C#.NET: Edit PDF Metadata.
add metadata to pdf file; endnote pdf metadata
C# PDF Print Library: Print PDF documents in C#.net, ASP.NET
view, Annotate,Convert documents online using ASPX. NET rotate PDF pages, C#.NET search text in PDF, C#.NET edit PDF bookmark, C#.NET edit PDF metadata, C#.NET
google search pdf metadata; pdf metadata viewer
FluoroMax
®
-4 & FluoroMax
®
-4P with USB v. B (23 Oct 2009) 
Troubleshooting 
7-11 
Open the Experiment Setup window: 
Click the Detectors icon 
Move the mouse over the detectors’ table in the Select area. 
The SpectrAcq firmware version appears in a small pop-up window: 
If the problem persists or is unlisted, call the Fluorescence Service De-
partment at (732) 494-8660 × 160. 
VB.NET PDF Print Library: Print PDF documents in vb.net, ASP.NET
view, Annotate,Convert documents online using ASPX. NET rotate PDF pages, C#.NET search text in PDF, C#.NET edit PDF bookmark, C#.NET edit PDF metadata, C#.NET
edit pdf metadata online; view pdf metadata in explorer
C# PDF File Compress Library: Compress reduce PDF size in C#.net
compressing control is designed to offer C# developers to compress existing PDF documents in .NET Document and metadata. C#.NET DLLs: Compress PDF Document.
online pdf metadata viewer; pdf metadata
FluoroMax
®
-4 & FluoroMax
®
-4P with USB v. B (23 Oct 2009) 
Troubleshooting 
7-12 
VB.NET PDF File Merge Library: Merge, append PDF files in vb.net
VB.NET Guide and Sample Codes to Merge PDF Documents in VB.NET Project. Batch merge PDF documents in Visual Basic .NET class program.
pdf metadata online; pdf metadata reader
VB.NET PDF Library SDK to view, edit, convert, process PDF file
for .NET empowers VB.NET developers to implement fast and high quality PDF conversions to or from multiple supported images and documents. PDF Metadata Edit.
embed metadata in pdf; pdf metadata editor
FluoroMax
®
-4 & FluoroMax
®
-4P with USB v. B (23 Oct 2009) 
Producing Correction Factors 
8-1 
8: Producing Correction Factors 
Introduction 
Gratings, detectors and other spectrometer components have response characteristics 
that are functions of wavelength. These characteristics are superimposed on spectra, 
and may yield a potentially misleading trace. For accurate intensity comparisons, such 
as those required for quantum-yield determinations, spectrometer-response characteris-
tics must be eliminated. Corrections are made for each of these potential problems by 
using radiometric correction factors. 
Supplied with the FluoroMax
®
-4 are sets of excitation and emission correction factors 
designed to eliminate response characteristics. These files, xcorrect.spc and 
mcorrect.spc, are included with the software and should be copied to the hard 
disk. The excitation correction range is from 240–600 nm, and the correction range for 
emission spectra is from 290–850 nm. 
C# PDF File Merge Library: Merge, append PDF files in C#.net, ASP.
C#.NET PDF Library - Merge PDF Documents in C#.NET. Provide NET components for batch combining PDF documents in C#.NET class. Powerful
pdf xmp metadata editor; read pdf metadata
C# TIFF: TIFF Metadata Editor, How to Write & Read TIFF Metadata
Our .NET SDKs own the most advanced & comprehensive documents and images reading and has a mature imaging utility which allows Tiff image file metadata to be
edit multiple pdf metadata; pdf keywords metadata
FluoroMax
®
-4 & FluoroMax
®
-4P with USB v. B (23 Oct 2009) 
Producing Correction Factors 
8-2 
Generating emission correction factors 
Generate a new correction-factor file only when the gratings or detectors have been re-
placed with those of different specifications than the original hardware. 
Required kits 
Emission correction factors should be updated periodically or whenever different grat-
ings or signal detectors are installed. The correction factors can be updated either at the 
user’s location, or by a representative from the Fluorescence Service Department. To 
arrange for a visit and a fee estimate, call the Fluorescence Service Department. To up-
date the correction factors without a service visit, call the Fluorescence Service De-
partment for instructions and further information. 
One way to generate correction factors for the instrument is to scan the spectrum of a 
standard lamp. Because the actual irradiance values of the standard lamp as a function 
of wavelength are known, dividing the irradiance values by the lamp spectrum results 
in a set of relative correction values. These values can then be applied to the raw fluo-
rescence data. The emission correction factor file mcorrect.spc was acquired in 
this manner. 
To generate emission correction factors, several items are needed: a standard lamp, ap-
propriate holders, and a light-scattering assembly. HORIBA Scientific offers the F-
3026 Standard Lamp Accessory. 
The F-3026 Standard Lamp Assembly is a complete correction-factor kit, which in-
cludes the following items: 
•  Calibrated lamp 
•  Irradiance values for the calibrated lamp 
•  Power supply 
•  Neutral-density filter 
•  Sample-compartment cover 
FluoroMax
®
-4 & FluoroMax
®
-4P with USB v. B (23 Oct 2009) 
Producing Correction Factors 
8-3 
Calculating emission correction factors 
Introduction 
For more information about the theory and application of radiometric correction, con-
sult Accuracy in Spectrophotometry and Luminescence Measurements, Mavrodineau, 
Schultz, and Menis, NBS Spec. Publ. 378 (1973), especially p. 137, “Absolute Spec-
trofluorometry,” by W.H. Melhuish. 
Irradiance values for a standard lamp, packaged with the lamp, usually are expressed in 
10
–6
W·cm
–2
·nm. With photon-counting systems like the FluoroMax
®
-4 spectrofluoro-
meter, however, data usually are collected in units of photons·s
–1
·cm
–2
·nm. To convert 
the units, multiply each irradiance value by the wavelength at which it is valid. (The da-
ta will still be off by a factor of c, but normalizing the correction factors compensates 
for this.) Such a mathematical procedure can be done in a spreadsheet program. 
Load the irradiance values. 
Enter the irradiance values into a spreadsheet. 
Load the irradiance file supplied by HORIBA Scientific by opening the 
FluorEssence™ software, choose File in the toolbar, choose Import, and 
then choose SPC file. 
Save the file as IRR. 
Now you have the two files: IRR and stdlamp2. These files are re-
quired to calculate the emission correction factors for the FluoroMax
®
-4. 
FluoroMax
®
-4 & FluoroMax
®
-4P with USB v. B (23 Oct 2009) 
Producing Correction Factors 
8-4 
Note:
Naming  the  file 
mcorrect  overwrites 
the  mcorrect file  sup-
plied with the software. 
Calculate the correction factors. 
Using the Simple Math Tool window 
under the Analysis/HJY Tools menu, 
divide IRR by stdlamp2, and name 
the resulting file mcorrect: 
stdlamp2
IRR
mcorrect =
Normalize the new mcorrect file. 
Using the HJY_normalize window under the Analysis/HJY Tools 
menu, divide the mcorrect file by this minimum signal intensity, a 
constant.  
Save this new file as mcorrect. 
(That is, overwrite the existing mcorrect file.) This normalizes the 
correction factor file so that the minimum intensity of mcorrect will 
be 1 count s
–1
. mcorrect contains the emission correction factors for 
the system. The correction-factor file should look similar to this: 
Once the emission correction factors are found, determination of the excitation correc-
tion factors may be necessary. The following procedures describe how to obtain excita-
tion correction factors using the photomultiplier and the photodiode. Follow the proce-
dure that applies to your configuration. 
FluoroMax
®
-4 & FluoroMax
®
-4P with USB v. B (23 Oct 2009) 
Producing Correction Factors 
8-5 
Note:
The  source/integrating-sphere  module  may  arrive  pre-
mounted atop the special gap-bed.
bed.
Calculating emission correction factors 
Set up the F-3026 accessory. 
Attach the correction-
factor light-source to its 
gap-bed with the brass 
thumbscrews.  
Use this orientation. Be sure the power 
cable doesn’t block optical apertures. 
Plug the power cable 
into the front of the gap-
bed.  
Remove the standard 
sample-compartment 
gap-bed, and insert the 
correction-factor kit 
gap-bed into the sam-
ple compartment. 
Fix the gap-bed firmly to the sample compartment with screws through the front 
panel. 
Run the scan.  
Start FluorEssence™ 
and initialize the instru-
ment. 
Turn the xenon lamp OFF, and the lamp 
for the correction-factor kit ON. 
ON 
FluoroMax
®
-4 & FluoroMax
®
-4P with USB v. B (23 Oct 2009) 
Producing Correction Factors 
8-6 
Note:
We recommend keeping a log of the amount of time the 
source is on. Calibration is necessary after every 50 h of use.
e.
Put the raised lid on the 
sample compartment.  
Make sure it fits securely, and there is no light 
leak into the sample compartment. 
Collect an emission scan using the following 
scan parameters: 
Range: 290–850 nm; increment: 5 nm; integration time: 1 s 
Warning:
Wear appropriate eye-
protection against UV, visible, 
and IR when the tungsten-
halogen lamp is on and the sam-
ple-compartment is uncovered. 
Caution:
Do not op-
erate the accessory 
without the lid com-
pletely covering the 
sample compart-
ment.