FluoroMax
®
-4 & FluoroMax
®
-4P with USB v. B (23 Oct 2009) 
Producing Correction Factors 
8-7 
Note:
Choose the correct irradiance file associated with the de-
sired detector. 
Analyze the data. 
Using the 
HJY_simple_math 
window under the 
Analysis/HJY 
Tools menu, di-
vide the irrad 
file by the 
stdlamp file:  
The irrad file is the irradiance file associated with the detector. 
The new graph now displays the correction factor. 
Normalize the new mcorrect file. 
Using the HJY_normalize window under the Analysis/HJY Tools 
menu, divide the irrad file by this minimum signal intensity, a 
constant.  
Pdf xmp metadata viewer - add, remove, update PDF metadata in C#.net, ASP.NET, MVC, Ajax, WinForms, WPF
Allow C# Developers to Read, Add, Edit, Update and Delete PDF Metadata
edit pdf metadata acrobat; adding metadata to pdf
Pdf xmp metadata viewer - VB.NET PDF metadata library: add, remove, update PDF metadata in vb.net, ASP.NET, MVC, Ajax, WinForms, WPF
Enable VB.NET Users to Read, Write, Edit, Delete and Update PDF Document Metadata
adding metadata to pdf files; pdf metadata editor online
FluoroMax
®
-4 & FluoroMax
®
-4P with USB v. B (23 Oct 2009) 
Producing Correction Factors 
8-8 
Warning:
Wear appropriate eye-
protection against UV, visible, and 
IR when the tungsten-halogen lamp 
is on and the sample-compartment 
is uncovered. 
Save this new file as irrad2. 
This normalizes the correction factor file so that the minimum intensity 
will be 1 count s
–1
. Highlight the B(Y) column, right-click and select 
Normalize. 
Export the normalized 
correction-factor 
graph.  
In the File menu, click Export, and 
select SPC Data…. 
Apply the correction 
factor into your sys-
tem configuration by 
following the instruc-
tions in your instru-
ment manual. 
Remove the accessory. 
Remove the sample-compartment cover.  
C# TIFF: TIFF Metadata Editor, How to Write & Read TIFF Metadata
How to Get TIFF XMP Metadata in C#.NET. Use this C# sample code to get Tiff image Xmp metadata for string. // Load your target Tiff docuemnt.
preview edit pdf metadata; remove pdf metadata online
DocImage SDK for .NET: Document Imaging Features
a metadata viewer application Enable users to add metadata in the form of EXIF, IPTC, XMP, or COM Type 6 (OJPEG) encoding Image only PDF encoding support.
read pdf metadata online; read pdf metadata java
FluoroMax
®
-4 & FluoroMax
®
-4P with USB v. B (23 Oct 2009) 
Producing Correction Factors 
8-9 
Caution:
The 
accessory 
may be 
HOT. Let 
the acces-
sory cool 
before re-
moval! 
Note:
We recommend keeping a log of the amount of time the 
source is on. Calibration is necessary after every 50 h of use.
e.
Turn OFF 
the acces-
sory.  
Remove the accessory’s 
sample-compartment 
gap-bed, and return the 
standard gap-bed to the 
sample compartment. 
XDoc.Tiff for .NET, Comprehensive .NET Tiff Imaging Features
types, including EXIF tags, IIM (IPTC), XMP data, and to read, write, delete, and update Tiff file metadata. Render and output text to text, PDF, or Word file.
remove metadata from pdf online; analyze pdf metadata
C# Raster - Raster Conversion & Rendering in C#.NET
RasterEdge XImage.Raster conversion toolkit for C#.NET supports image conversion between various images, like Jpeg, Png, Bmp, Xmp and Gif, .NET Graphics
remove metadata from pdf file; batch pdf metadata
oolbar, 
C# Raster - Image Process in C#.NET
Image Access and Modify. Image Information. Metadata(tag) Edit. Color VB.NET How-to, VB.NET PDF, VB.NET Word process various images, like Jpeg, Png, Bmp, Xmp
acrobat pdf additional metadata; add metadata to pdf programmatically
.NET JPEG 2000 SDK | Encode & Decode JPEG 2000 Images
Home > .NET Image Viewer > jpeg 2000. Able to customize compression ratios (0 - 100); Support metadata encoding and decoding, including IPTC, XMP, XML Box
pdf xmp metadata; bulk edit pdf metadata
FluoroMax
®
-4 & FluoroMax
®
-4P with USB v. B (23 Oct 2009) 
Producing Correction Factors 
8-11 
Note:
The  signal  detector,  S,  uses  only 
the  mcorrect  file;  the  reference  detec-
tor, R, uses only the xcorrect file. 
The Preferences area appears. 
Choose the Instrument Correction Files icon. 
The Instrument Correction Files area appears. 
If there are no active fields in the Instrument 
Correction Files area, click the Insert button. 
In sequence, 
Choose the 
Detector 
column, and 
select the appropriate detector from the drop-down list, 
The Mono column, and the correct monochromator from the drop-down 
list, 
The Grating column, the correct grating from the drop-down list, 
Then, in the File column, browse for the appropriate correction-factor 
file. 
FluoroMax
®
-4 & FluoroMax
®
-4P with USB v. B (23 Oct 2009) 
Producing Correction Factors 
8-12 
When all necessary detectors have an asso-
ciated correction-factor file, click the OK button. 
Reinitialize the instrument and software. 
In the FluorEssence™ toolbar, choose Collect. 
In the drop-down menu, choose Advanced Setup. 
In the drop-down menu, choose System ReInitialization. 
The new correction-factor files are now ready to be activated in the Experiment 
Setup window, under the Detectors icon: 
FluoroMax
®
-4 & FluoroMax
®
-4P with USB v. B (23 Oct 2009) 
Producing Correction Factors 
8-13 
Excitation correction factors 
Excitation correction factors are measured during production of the instrument. If you 
have further questions, please contact the Fluorescence Service Department.
FluoroMax
®
-4 & FluoroMax
®
-4P with USB v. B (23 Oct 2009) 
Producing Correction Factors 
8-14 
FluoroMax
®
-4 & FluoroMax
®
-4P with USB v. B (23 Oct 2009) 
FluoroMax
®
-4P Phosphorimeter Operation 
9-1 
9: FluoroMax
®
-4P Phosphori-
meter Operation 
Introduction 
The FluoroMax
®
-4P includes a phosphorimeter, that is, a programmable pulsed source 
and selectable signal gating from the reference detector. Switching between the pulsed 
lamp and continuous lamp is computer-controlled. Apart from this, the FluoroMax
®
-4P 
is identical in operation in all other respects to the FluoroMax
®
-4. 
FluoroMax
®
-4 & FluoroMax
®
-4P with USB v. B (23 Oct 2009) 
FluoroMax
®
-4P Phosphorimeter Operation 
9-2 
Theory of operation 
A second source of illumination, a pulsed xenon lamp, is used for phosphorescence 
measurements. Samples are excited with pulsed light; the emitted phosphorescence is 
measured using an R928P photon-counting detector. 
Sequence of data acquisition 
The illuminator housing, or flash lamp, operates at up to 25 Hz. The control module 
triggers each lamp pulse. When the start of the light output is detected, a signal is sent 
to the control module for timing purposes. The control module houses the signal-gating 
circuitry that intercepts the signal from the pulse-counting emission photomultiplier 
tube, collects a selected, time-delimited portion of the signal, and later passes it to the 
software. The maximum signal detectable per flash varies with the integration time: 
Integration time  Maximum signal (counts) per flash 
100 μs 
400 
1 ms 
4000 
10 ms 
40 000 
100 ms 
400 000 
1 s 
4 000 000 
With, for example, 100 flashes of integrated over 10 ms each, the maximum detectable 
signal is: 
100 flashes × 40 000 counts = 4 000 000 counts total 
A typical sequence of data-acquisition (see below) starts with a flash from the pulsed 
lamp, sensed by the control module as time t = 0. The light enters the excitation mo-
nochromator, where it is dispersed. Monochromatic light from the monochromator ex-
cites the sample. Luminescence emission from the sample then passes through the 
emission monochromator to the photomultiplier-tube detector. The control module in-
cludes a gate-and-delay generator, allowing the signal at the detector to be integrated 
only during a specific period after the flash (the Initial Delay), for a pre-determined 
length of sampling time (the Sample Window). Any signal arriving before or after the 
gating is ignored. 