FluoroMax
®
-4 & FluoroMax
®
-4P with USB v. B (23 Oct 2009) 
FluoroMax
®
-4P Phosphorimeter Operation 
9-3 
This sequence of excitation, delay, and sampling, is 
repeated for each lamp flash. The signal is accumu-
lated for a predetermined number of excitation 
pulses, then FluorEssence™ collects the total signal. 
After collection, FluorEssence™ displays the inten-
sity of the luminescence as a function of time or wa-
velength. The x-axis is based on one of the four scan 
options: 
•  Decay by Delay 
•  Decay by Window 
•  Emission 
•  Excitation 
Phosphorimeter parameters 
Four FluorEssence™ parameters govern the sequence in a phosphorimeter experiment. 
These parameters automatically appear on the phosphorimeter experiment-acquisition 
window. 
Initial De-
lay 
Sets the time, in ms, between the start of the lamp flash and the onset of 
data-acquisition (opening of the Sample Window). Initial Delay can 
range from 0–10 000 ms, in increments of 0.001 ms. Accuracy of Initial 
Delay is better than ± 0.001 ms. 
Set Initial Delay long enough so that fluorescence emission and lamp de-
cay are complete, so that the resulting spectrum represents phosphores-
cence only. The full-width at half-maximum lamp-pulse width is 3 μs, but 
there is a long decay time for the light output. Lamp intensity falls to less 
than 1% of peak output after 45 μs. Setting the delay to > 50 μs effectively 
Batch edit pdf metadata - add, remove, update PDF metadata in C#.net, ASP.NET, MVC, Ajax, WinForms, WPF
Allow C# Developers to Read, Add, Edit, Update and Delete PDF Metadata
view pdf metadata; pdf metadata extract
Batch edit pdf metadata - VB.NET PDF metadata library: add, remove, update PDF metadata in vb.net, ASP.NET, MVC, Ajax, WinForms, WPF
Enable VB.NET Users to Read, Write, Edit, Delete and Update PDF Document Metadata
pdf xmp metadata viewer; delete metadata from pdf
FluoroMax
®
-4 & FluoroMax
®
-4P with USB v. B (23 Oct 2009) 
FluoroMax
®
-4P Phosphorimeter Operation 
9-4 
removes any interference from the lamp. 
Initial Delay can be varied with time to yield a decay curve. Spectra can 
be scanned to isolate different phosphorescing components based on the 
lifetime of the luminescent species in the sample. Together, these two 
techniques can be used to create three-dimensional plots. For example, 
successive scans with varying delay times can be plotted. 
To record fluorescence and phosphorescence emission, set Initial Delay to 
zero. 
Sample 
Window 
Sets the duration of signal acquisition, in ms. The Sample Window opens 
when the Initial Delay ends. When the Sample Window opens, the signal 
is counted and integrated. After the Sample Window closes, any signal is 
ignored. 
The Sample Window may be set from 0.01 to 10 000 ms. If the lifetime 
of the phosphorescence is known, set the Sample Window to 5–10 times 
the lifetime. If the phosphorescence lifetime is unknown, make the Sam-
ple Window a small fraction of the anticipated lifetime, and then increase 
it until acceptable results are observed. 
If the Sample Window is too long, the detector will record spurious 
background signal. If the Sample Window is too short, components of 
the lifetime decay may be missed. With two or more species decaying si-
multaneously, try varying the Initial Delay and the Sample Window. 
Time per 
flash 
Sets the total cycle length per flash, including on time, decay time, and 
dead time between flashes. The Time per flash is the reciprocal of the 
repetition rate of the lamp pulses. The allowable repetition rate is 0.03–25 
Hz. The Time per flash must be slow enough to let the Sample Window 
close before another flash begins. Accuracy of the repetition rate is ± 1 ms. 
The Time per flash is governed by 
Time per Flash ≥ Initial Delay + Sample Window + 20 ms 
Flash 
count 
Sets the number of lamp pulses that contribute to each data point. The 
range for Flash count is 1 to 999. The signal collected per flash is inte-
grated over the total Flash counts before FluorEssence™ stores it. The 
more flashes accumulated, the higher the signal-to-noise ratio becomes. 
For more information about FluorEssence™ phosphorimeter commands, consult the 
FluorEssence™ on-line help. 
C# PDF Convert to Tiff SDK: Convert PDF to tiff images in C#.net
Studio .NET project. Powerful .NET control to batch convert PDF documents to tiff format in Visual C# .NET program. Free library are
pdf metadata viewer online; metadata in pdf documents
C# PDF Convert to Word SDK: Convert PDF to Word library in C#.net
Powerful components for batch converting PDF documents in C#.NET program. In the daily-life applications, you often need to use and edit PDF document content
c# read pdf metadata; search pdf metadata
FluoroMax
®
-4 & FluoroMax
®
-4P with USB v. B (23 Oct 2009) 
FluoroMax
®
-4P Phosphorimeter Operation 
9-5 
Applications for the phosphorimeter 
Phosphorescence decay curve 
With the Phos experiment type in the Fluorescence Experiment Menu, create a phos-
phorescence-decay curve, as in the screenshot from FluorEssence™ shown above. This 
is an example using data from TbCl
3
(aq), fitted to a single-exponential curve-fit in Ori-
gin
®
. All parameters are automatically displayed, including the fitted lifetime of 427 μs. 
VB.NET Create PDF from Excel Library to convert xlsx, xls to PDF
NET edit PDF metadata, C#.NET edit PDF digital signatures, C#.NET edit PDF sticky note Professional .NET PDF converter component for batch conversion.
batch edit pdf metadata; batch pdf metadata editor
VB.NET PDF Convert to Jpeg SDK: Convert PDF to JPEG images in vb.
C#.NET edit PDF digital signatures, C#.NET edit PDF sticky note Best and free VB.NET PDF to jpeg converter SDK for Visual NET components to batch convert adobe
rename pdf files from metadata; pdf remove metadata
FluoroMax
®
-4 & FluoroMax
®
-4P with USB v. B (23 Oct 2009) 
FluoroMax
®
-4P Phosphorimeter Operation 
9-6 
Isolate components in a mixture based on lifetimes 
To the right are three scans 
of an aqueous mixture of 
terbium and europium chlo-
rides that isolate different 
phosphorescent components 
based on their lifetimes. For 
example, a sample contain-
ing two phosphorescent 
species with different life-
times (Tb = 421 μs; Eu = 
113 μs) can be resolved into 
its components by varying 
the Initial Delay parameter, 
here shown as bold italic 
numerals, 0.1 ms, 1 ms, and 
2 ms initial delay. Notice 
the wavelength shift in the 
593 nm peak (marked with 
a gray line) from initially 
mostly europium fluorescence (short lifetime) to a mixture of terbium (longer lifetime) 
and europium fluorescence at later delays. Also notice the Eu peak near 690 nm at 0.1 
ms that vanishes at later times. This experiment was performed using the Emission 
subtype of Phos experiment. 
Three-dimensional plots 
Using the phosphorescence delay curve with the Initial Delay technique, a three-
dimensional graph can be created, as shown below. This graph shows successive scans 
of a mixture of Tb-L and Eu-L (where L = benzophenone antenna chromophore) with 
varying time delays, segregating species into contours representing a particular slice of 
time. 
C# PDF File Merge Library: Merge, append PDF files in C#.net, ASP.
NET components for batch combining PDF documents in C#.NET class. Powerful library dlls for mering PDF in both C#.NET WinForms and ASP.NET WebForms.
change pdf metadata; extract pdf metadata
VB.NET PDF Convert to Tiff SDK: Convert PDF to tiff images in vb.
NET control to batch convert PDF documents to Tiff format in Visual Basic. Qualified Tiff files are exported with high resolution in VB.NET.
batch update pdf metadata; remove metadata from pdf
FluoroMax
®
-4 & FluoroMax
®
-4P with USB v. B (23 Oct 2009) 
FluoroMax
®
-4P Phosphorimeter Operation 
9-7 
Kinetic analysis of mixtures 
Often a sample containing a mixture of components can be analyzed through fitting its 
phosphorescence-decay curve. Here is a phosphorescence decay of an aqueous mixture 
of EuCl
3
and TbCl
3
, whose different lifetimes have been extracted by FluorEssence™’s 
dual-exponential fit. Curve-fitting merely involves choosing the analytical model. 
VB.NET PDF Convert to Word SDK: Convert PDF to Word library in vb.
project. Professional .NET library supports batch conversion in VB.NET. .NET control to export Word from multiple PDF files in VB.
remove metadata from pdf acrobat; edit pdf metadata
C# PDF Print Library: Print PDF documents in C#.net, ASP.NET
C#.NET edit PDF metadata, C#.NET edit PDF digital signatures, C#.NET edit PDF sticky note, C# Quicken PDF printer library allows C# users to batch print PDF
remove pdf metadata; change pdf metadata creation date
FluoroMax
®
-4 & FluoroMax
®
-4P with USB v. B (23 Oct 2009) 
FluoroMax
®
-4P Phosphorimeter Operation 
9-8 
Operation of the phosphorimeter 
Start-up 
Load an appropriate instrument configuration that includes the phosphorimeter. 
FluorEssence™ features 
Main Experiment menu 
This is where you choose to scan with the 
phosphorimeter, via a Phos scan: 
Experiment Type window 
After choosing the Phos experiment type in the Fluorescence Main Experiment Menu, 
four experiment types are available for the phosphorimeter in the Experiment Type 
window:  
Excitation 
Emission 
These are similar to the standard excitation, and emission scans. Among 
the changes are that the Integration Time field is removed, and instead 
are data-entry fields for Sample window, Delay after flash, Time per 
flash, and Flash count. These four fields are described above, in the 
Theory of Operation section, and shown below (extracted from the Expe-
riment Setup window). 
FluoroMax
®
-4 & FluoroMax
®
-4P with USB v. B (23 Oct 2009) 
FluoroMax
®
-4P Phosphorimeter Operation 
9-9 
Above is a Phosphorimeter area with a typical set of parameters for the emission of 
EuCl
3
. Use an excitation monochromator set to 393 nm; the emission monochromator 
should start at 570 nm, end at 750 nm, with an increment of 1 nm. 
Decay by 
Delay 
Decay by 
Window 
These produce a decay of phosphorescence over time. Decay by Delay 
varies the flash’s Delay after flash in order to construct the decay curve. 
Decay by Window varies the length of the Sample window while tak-
ing data. 
Above is a Phosphorimeter area showing typical parameters for the phosphorimeter 
decay of EuCl
3
. Set the emission monochromator to 590 nm and the excitation monoch-
romator to 393 nm. 
Experiment Setup window 
The Experiment Setup window, under the Detectors icon, includes a Phosphorimeter 
area, indicating that the phosphorimeter is available. 
FluoroMax
®
-4 & FluoroMax
®
-4P with USB v. B (23 Oct 2009) 
FluoroMax
®
-4P Phosphorimeter Operation 
9-10 
Processing phosphorimeter data 
Open the graph to be processed. 
Click on the data points to be processed. 
In the tool-
bar, choose 
Analysis. 
A drop-down menu 
appears. 
Choose Fit. 
From the 
submenu, 
choose the 
type of ana-
lytical curve 
to use. 
Follow the 
instructions 
for that type 
of curve. 
The fit plus parameters appear on the graph. 
FluoroMax
®
-4 & FluoroMax
®
-4P with USB v. B (23 Oct 2009) 
FluoroMax
®
-4P Phosphorimeter Operation 
9-11 
Warning:
High voltage exists within the FluoroMax
®
-4P. 
To avoid fatal shocks, before removing the lamp cover, 
unplug the FluoroMax
®
-4P’s power cord, and wait at 
least 5 min while the internal capacitors discharge. 
Never operate the lamp with the cover removed! 
Warning:
Xenon lamps are an explosion hazard. Be 
sure that the power is off, and all AC (mains) power is 
disconnected from the system. Read and follow all the 
cautions below: 
Lamp replacement 
The xenon flash lamp typically has a half-intensity life of at least 10 million flashes. 
Follow the procedure below for replacement and alignment. 
FluoroMax
®
-4 & FluoroMax
®
-4P with USB v. B (23 Oct 2009) 
FluoroMax
®
-4P Phosphorimeter Operation 
9-12 
Caution:
Never rotate the flip mirror inside the 
lamp housing manually. This can strip the 
gears in the gearbox.
Hazards 
Xenon-arc lamps are an explosion hazard. Wear explosion-proof face-shield and pro-
tective clothing when opening the lamp housing and handling the lamp. 
Disconnect the lamp power supply from the AC power line (mains) while handling 
lamp leads. Lethal high voltages may exist. 
The lamp remains extremely hot for approximately one-half hour after it has been 
turned off. Do not touch the lamp or the metal unit until the lamp has cooled. 
Never look directly at the xenon arc or its reflection. Intense radiation can permanently 
damage eyes. 
Do not touch the focusing lens, back-scatter mirror, or the surface of the lamp. Finger-
prints will be burned onto these surfaces when the lamp is ignited. 
Method 
In FluorEssence™, be sure the FluoroMax
®
-4P 
instrument configuration is loaded. 
The flip-mirror automatically rotates to the flash lamp, giving you more room to 
work. 
Switch off and prepare the FluoroMax
®
-4P. 
Be sure that the FluoroMax
®
-4P and the host computer are turned off. 
Remove the AC (mains) power cord from the FluoroMax
®
-4P. 