R&S ZVL 
GUI Reference 
Channel Menu 
Operating Manual 1303.6580.32-06 
235
Remote 
control: 
CALCulate<Ch>:TRANsform:VNETwork:SENDed... 
MMEMory:LOAD:VNETworks<Ch>:SENDed:DEEMbedding<Ph_pt> 
'<file_name>' 
MMEMory:LOAD:VNETworks<Ch>:SENDed:EMBedding<Ph_pt> '<file_name>'
Cal Manager     
Opens a dialog to store system error correction and power correction data to the cal pool and to assign 
stored correction data to channels.   
Cal pool 
The cal pool is a collection of correction data sets (cal groups) that the analyzer stores in a common 
directory C:\Rohde&Schwarz\NWA\Calibration\Data. Cal groups in the pool can be applied to different 
channels and setups. Each cal group is stored in a separate file named <CalGroup_name>.cal. The cal 
group name can be changed in the Calibration Manager dialog.  
The Calibration Manager dialog contains three main panels:  
Calibration State lists all channels and their current calibration state.  
Calibration Pool contains a list of all stored correction data sets.  
Calibration Properties displays the basic channel settings, the System Error Correction Data, and 
Remove metadata from pdf file - add, remove, update PDF metadata in C#.net, ASP.NET, MVC, Ajax, WinForms, WPF
Allow C# Developers to Read, Add, Edit, Update and Delete PDF Metadata
pdf keywords metadata; endnote pdf metadata
Remove metadata from pdf file - VB.NET PDF metadata library: add, remove, update PDF metadata in vb.net, ASP.NET, MVC, Ajax, WinForms, WPF
Enable VB.NET Users to Read, Write, Edit, Delete and Update PDF Document Metadata
pdf xmp metadata; online pdf metadata viewer
R&S ZVL 
GUI Reference 
Channel Menu 
Operating Manual 1303.6580.32-06 
236
the Power Correction Data available for the channel selection in the Calibration State panel.  
Calibration states 
One of the following calibration states is assigned to each of the channels listed in the Cal State table:  
No Calibration No specific channel correction data is available. The factory system error correction 
is used. This is the default situation when a new setup is created (Nwa-File – New).  
Channel Cal A system error correction and/or power correction with channel-specific correction 
data is active. A new calibration of the channel affects the channel-specific correction data only.    
Uses <CalGroup> A system error correction with one of the correction data sets (cal groups) 
stored in the cal pool is active. A new calibration of the channel replaces the correction data in the 
cal pool and may affect several channels, see below.  
The channel calibration of the active channel is also applied to a new channel created via New Add Chan 
+Trace indicates that the system error correction for the channel has been deactivated using Correction 
Off. 
If a new calibration is performed for a channel assigned to a cal group (marked as Uses <CalGroup>), 
the correction data overwrites the cal group data, so the new calibration will affect all channels 
assigned to the cal group. The network analyzer generates a notice message "New calibration will 
overwrite cal pool!" when opening the first dialog of the calibration wizard. 
The buttons in the upper half of the dialog are used to modify the Cal Pool and apply pool data to 
channels: 
Copy >> copies the correction data (Channel Cal) of the selected channel to the cal pool, 
generating a new pool member (cal group).   
Apply assigns the selected cal group to the selected channel.  
Apply All assigns the selected cal group to all channels in the Cal State table.  
Resolve Pool Link deletes the assignment between the selected channel (Ch<n>: Uses 
<CalGroup_name>) and the cal group. The correction data from the cal group <CalGroup_name> 
is used as a channel calibration for <Ch1>, the Calibration State list displays Ch<n>: Channel Cal. 
Delete From Pool deletes the selected cal group. 
You can apply any correction data set (cal group) from the Cal Pool to several channels, which may 
belong to different setups. When doing make sure that the port configuration of the channels 
(connectors, state, ...) and the cal group are compatible.  
You can also use a cal group stored in the Cal Pool to define a channel-specific calibration. Apply 
the cal group to the channel and click Resolve Pool Link.   
Remote 
control: 
MMEMory:STORe:CORRection <Ch>, "<file_name>" 
MMEMory:LOAD:CORRection <Ch>, "<file_name>" 
MMEMory:LOAD:CORRection:RESolve <Ch>, "<file_name>" 
MMEMory:DELete 
"C:\R_S\Instr\user\Nwa\Calibration\Data\<file_name>" 
MMEMory:DELete:CORRection "<file_name>"  
[SENSe<Ch>:]CORRection:DATE? 
[SENSe<Ch>:]CORRection:DATA:PARameter? 
[SENSe<Ch>:]CORRection:SSTate? 
Recall Last Cal Set 
Loads and activates the setup for which the last calibration was performed. If the last calibrated 
C# PDF Password Library: add, remove, edit PDF file password in C#
C# Sample Code: Remove Password from PDF File in C#.NET. These C# demos will help you to delete password for an encrypted PDF file. // Define input file path.
pdf metadata reader; remove metadata from pdf acrobat
VB.NET PDF File Compress Library: Compress reduce PDF size in vb.
Remove bookmarks, annotations, watermark, page labels and article threads from PDF Document and metadata. NET Demo Code to Optimize An Exist PDF File in Visual
pdf metadata viewer; batch update pdf metadata
R&S ZVL 
GUI Reference 
Channel Menu 
Operating Manual 1303.6580.32-06 
237
setup is already active, nothing is changed. 
The calibrated setups are automatically stored in the 
C:\R_S\Instr\user\Nwa\Calibration\RecallSets directory. A message box pops up if the directory is 
empty, e.g. because no calibration was performed yet.  
Automatic Calibration (Procedure)     
A Calibration  Unit  is  an  integrated  solution  for  automatic  system  error  correction  of  vector  network 
analyzers. For an overview of available types refer to section Automatic Calibration (Introduction). The use 
of the calibration units is described below. 
Attention! 
Please  observe  the  safety  instructions  in  the  "Technical  Information" 
provided with the calibration unit to avoid any damage to the unit and the 
network  analyzer.  Safety-related  aspects  of  the  connection  and 
operation of the units are also reported in the sections below.   
Connecting the Calibration Unit to the Network Analyzer 
The calibration unit provides two N 75 V (f) connectors.  
To connect the unit... 
Switch on and start-up your network analyzer.  
To protect your equipment against ESD damage use the wrist strap and grounding cord supplied with 
the instrument and connect yourself to ground. GND 
Connect the USB type A connector of the USB cable to any of the USB type A connectors on the front 
panel of  the analyzer.  Connect the USB type  B connector  of  the USB cable to  the  USB  type  B 
connector of the calibration unit. You can also connect the unit before switching on the analyzer.  
Wait until the operating system has recognized and initialized the new hardware. When the unit is 
connected for the first time, this may take longer than in normal use.   
The unit is ready to be used, see Performing an Automatic Calibration below.  
VB.NET PDF Page Delete Library: remove PDF pages in vb.net, ASP.
Ability to remove consecutive pages from PDF file in VB.NET. Enable specified pages deleting from PDF in Visual Basic .NET class.
edit pdf metadata online; preview edit pdf metadata
VB.NET PDF remove image library: remove, delete images from PDF in
Image: Insert Image to PDF. Image: Remove Image from PDF Page. Image Link: Edit URL. Bookmark: Edit Bookmark. Metadata: Edit, Delete Metadata. Form Process.
pdf metadata viewer online; edit pdf metadata acrobat
R&S ZVL 
GUI Reference 
Channel Menu 
Operating Manual 1303.6580.32-06 
238
Note:  
The  calibration  unit  is  intended  for  direct  connection  to  ZVB  network 
analyzers following the procedure described above. You can also connect the 
unit before switching on the analyzer. Do not connect the unit to other USB 
hosts, e.g. a PC, or insert any USB hubs between the analyzer and the unit, 
as this may cause damage to the unit or the host.  
You  can  connect  only  one  calibration  unit  at  a  time.  However,  you  can 
simultaneously connect a cal unit and another device (mouse, USB memory 
stick etc.) to the USB connectors of the analyzer.   
An unused calibration unit may remain connected to the USB port while the 
network  analyzer  is  performing  measurements.  It  must  be  disconnected 
during a firmware update.  
It  is  safe  to  connect  or  disconnect  the  calibration  unit  while  the  network 
analyzer  is  operating.  Never connect  or  disconnect  the unit  while  data  is 
being transferred between the analyzer and the unit. Never connect the unit 
during a firmware update.  
Performing an Automatic Calibration 
After connection and initialization, you can use the calibration unit as follows:   
Connect the analyzer ports  you  wish  to calibrate (one/two  ports  for a one/two-port  calibration) to 
arbitrary ports of the calibration unit. If you connect just one cal unit port, terminate the other port with 
a50 V match.  
Perform the automatic calibration for the selected number of ports using the S-Parameter Wizard or 
the Start Calibration menu. 
Remove  the  test  cables  from  the  unit,  connect  your  DUT  instead  and  perform  calibrated 
measurements.  
The assignment between the analyzer ports and the cal unit ports is detected automatically. If auto-
detection fails (e.g. because of a high attenuation in the signal path), connect matching port numbers and 
click Default Port Assignment at the beginning of the calibration process.  
Accuracy considerations 
To ensure an accurate calibration, please observe the following items:  
Unused ports of the calibration unit must be terminated with a 50 V match.  
No adaptors must be inserted between the calibration unit and the test ports.  
After  connecting  the  unit  to  the  USB  port,  allow  for  a  sufficient  warm-up  time  (see 
"Specifications") before starting the calibration.  
Highest accuracy is achieved if the input power at the cal unit ports is approx. –10 dBm. This 
corresponds to the source power of the R&S ZVL after a preset of the instrument. If the test 
setup  contains  a  large  attenuation,  increase  the  source  power  to  compensate  for  the 
attenuation.  
C# PDF Page Delete Library: remove PDF pages in C#.net, ASP.NET
Ability to remove a range of pages from PDF file. Free trial package for quick integration in .NET as well as compatible with 32 bit and 64 bit windows system.
metadata in pdf documents; pdf metadata
C# PDF File Compress Library: Compress reduce PDF size in C#.net
Remove bookmarks, annotations, watermark, page labels and article threads from PDF while compressing. Document and metadata. C#.NET DLLs: Compress PDF Document.
change pdf metadata; search pdf metadata
R&S ZVL 
GUI Reference 
Channel Menu 
Operating Manual 1303.6580.32-06 
239
Attention!  
The maximum RF input power of the calibration unit is beyond the 
RF output power range of the analyzer, so there is no risk of 
damage if the device is directly connected to the test ports. If you 
use an external power amplifier, make sure that the maximum RF 
input power of the calibration unit quoted in the data sheet is never 
exceeded.   
The calibration type depends on the number of ports. If a single port is calibrated, the analyzer uses full 
one-port calibration. For 2 ports, the analyzer provides the following options:  
ATOSM calibration for the two ports.  
One or two full one-port calibrations.  
Remote 
control
:
[SENSe<Ch>:]CORRection:COLLect:AUTO 
[SENSe<Ch>:]CORRection:COLLect:AUTO:CKIT 
[SENSe<Ch>:]CORRection:COLLect:AUTO:PORTs  
[SENSe<Ch>:]CORRection:COLLect:AUTO:PORTs:CONNection? 
Calibration Procedure 
The  automatic  calibration  is  started  on  clicking  Calibration  Unit  in  the  Start  Calibration  menu.  The 
Calibration Unit dialog shows the parameters of the cal unit, the port assignment, and the progress of the 
calibration procedure. 
In the dialog, you can:  
Detect the current Port Assignment between the analyzer and the connected calibration unit.  
Restore the Default Port Assignment (port 1 / 2 of the analyzer to port 1 / 2 of the cal unit). This is 
important if the automatic detection of the port assignment fails.  
Start the necessary calibration sweeps. The calibration is started automatically if automatic port 
detection is successful and the cal unit temperature is within the validity range.  
VB.NET PDF File Permission Library: add, remove, update PDF file
Image: Insert Image to PDF. Image: Remove Image from PDF Page. Image Link: Edit URL. Bookmark: Edit Bookmark. Metadata: Edit, Delete Metadata. Form Process.
embed metadata in pdf; c# read pdf metadata
C# PDF bookmark Library: add, remove, update PDF bookmarks in C#.
Help to add or insert bookmark and outline into PDF file in .NET framework. Ability to remove and delete bookmark and outline from PDF document.
read pdf metadata; preview edit pdf metadata
R&S ZVL 
GUI Reference 
Channel Menu 
Operating Manual 1303.6580.32-06 
240
Restore the Default Port Assignment (port 1 / 2 of the analyzer to port 1 / 2 of the cal unit). This is 
important if the automatic detection of the port assignment fails.  
Start the necessary calibration sweeps. The calibration is started automatically if automatic port 
detection is successful and the cal unit temperature is within the validity range.  
Abort an ongoing calibration sweep.  
The Calibration Unit dialog is closed automatically when all standards have been measured, the system 
error correction data has been calculated, and the result is applied to the active channel. The network 
analyzer is ready to perform calibrated measurements.  
If you accidentally connect a calibration unit which is no suited for R&S ZVL network analyzers 
(e.g. one of the cal unit types for R&S ZVA/B analyzers) the analyzer generates an error message, 
stating that the characterization of the cal unit is not sufficient.  
Remote 
control
:
[SENSe<Ch>:]CORRection:COLLect:AUTO 
[SENSe<Ch>:]CORRection:COLLect:AUTO:CKIT 
[SENSe<Ch>:]CORRection:COLLect:AUTO:PORTs  
[SENSe<Ch>:]CORRection:COLLect:AUTO:PORTs:CONNection? 
SYSTem:COMMunicate:RDEVice:AKAL:ADDRess  
SYSTem:COMMunicate:RDEVice:AKAL:ADDRess:ALL? 
SYSTem:COMMunicate:AKAL:CONNection  
SYSTem:COMMunicate:AKAL:MMEMory[:STATe] 
MMEMory:AKAL:FACTory:CONVersion 
Cal Kits     
Opens a dialog to manage the calibration kits in use, add new kits and import or export kits.  
Calibration kits 
Acalibration kit is a set of physical calibration standards for a particular connector type. The magnitude 
and phase response of the calibration standards (i.e. their S-parameters) must be known or predictable 
within a given frequency range.  
The standards are grouped into several types (open, through, match,...) corresponding to the different 
input quantities for the analyzer's error models. The standard type also determines the equivalent circuit 
model used to describe its properties. The circuit model depends on several parameters that are stored in 
the cal kit file associated with the calibration kit.  
As an alternative to using circuit models, it is possible to describe the standards by means of S-parameter 
tables stored in a file. 
The analyzer provides a large number of predefined cal kits but can also import cal kit files and create new 
kits:  
Aselection of predefined kits with different types of parameters is available for all connector types. 
The parameters of these kits are displayed in the Add/Modify Standards dialog, however, it is not 
possible to change or delete the kits.  
Imported and user-defined kits can be changed in the Calibration Kits dialog and its various sub-
dialogs.  
Calibration kits and connector types are global resources; the parameters are stored independently and 
available irrespective of the current setup.  
R&S ZVL 
GUI Reference 
Channel Menu 
Operating Manual 1303.6580.32-06 
241
The Calibration Kits dialog contains a series of buttons, each of them opening a dialog: 
Import Kit imports a cal kit file containing the parameters for a new calibration kit. It is possible to 
load several kits, so the analyzer asks whether to set the imported calibration kit active.  
Add Kit combines different standards to form a new calibration kit.   
View/ Modify Kit adds or deletes standards in an imported or user-defined calibration kit and/or 
changes their parameters.  
Copy Kit to... creates a copy of the selected calibration kit. The cal kit data is stored in the internal 
memory and can be exported to a cal kit file using Export Kit. 
Delete Kit removes an imported or user-defined kit from the list of Calibration Kits in Use.
Export Kit exports calibration kit data to a cal kit file.   
Avail Conn Types adds or deletes connector types.  
The Calibration Kits in Use: table is used to select a calibration kit for each connector type. The pull-down 
lists contain all kits for the individual connector types.   
Remote control:
[SENSe<Ch>:]CORRection:CKIT:<conn_type>:SELect '<ckit_name>' 
[SENSe<Ch>:]CORRection:CKIT:SELect '<conn_type>', '<ckit_name>'
Cal Kit Parameter Types  
The analyzer uses three types of parameters to describe the calibration standards. The parameter type is 
the same for all standards in a kit and therefore annexed to the kit name:  
Universal parameters (no label)  describe calibration kit models with highly standardized 
components so that the parameters are valid for all calibration kits of the model.  
Typical parameters (labelled typical) approximately describe a calibration kit model. To correct for 
deviations between the standards, each kit of the model is individually measured and delivered 
with an additional, kit-specific parameter set. Therefore each typical parameter set <kit_name> 
typical is complemented by an additional parameter set <kit_name> containing optimized 
parameters for an individual kit.  
Ideal parameters (labelled Ideal Kit) describe an idealized calibration kit for each connector type; 
R&S ZVL 
GUI Reference 
Channel Menu 
Operating Manual 1303.6580.32-06 
242
see below.  
Make sure to use universal or individual parameter sets if you need to obtain high-precision results. 
The precision of the calibration kit parameters determine the accuracy of the system error correction and 
of the measurements. The calibration wizard displays a warning if you use a typical or ideal parameter set 
to calibrate a channel.  
Calibration kits can be obtained as network analyzer accessories; refer to the data sheet for the relevant 
ordering information. The name of all parameter sets is equal to the name of the corresponding calibration 
kit model.  
Ideal parameters 
All ideal kits contain the following standards:  
Standard (Gender) 
R(Load) 
Electrical Length (Offset) 
Open (f, m) 
V
0mm (Delay: 0 s) 
Short (f, m) 
0V
0mm 
Offset Short (f, m) 
0V
10 mm 
Match (f, m) 
Z
0
(reference impedance of the connector type) 
0 mm 
Sliding Match (f, m) 
– 
0 mm 
Reflect (f, m) 
V
0mm 
Through (ff, mm, mf) 
– 
0 mm 
Line (ff, mm, mf) 
– 
10 mm 
Attenuation (ff, mm , mf) 
– 
0 mm 
Symm. Network (ff, mm, mf) 
– 
0 mm 
The following additional parameters are used:  
Characteristic impedance: Z
0
(reference impedance of the connector type) 
Loss: 0 dB / sqrt(GHz) (0 GV / s)  
All inductance and capacitance parameters are set to zero.  
Cal Kit Files  
Calibration kit files can be used to store the parameters of a particular calibration kit, to reuse the data and 
to exchange calibration kits from one network analyzer to another.  
Cal kit file contents 
Cal kit files are independent of the current setup and contain the following information:  
Name and label of the calibration kit 
Connector type including all connector type parameters (name, polarity, offset model, reference 
impedance)  
Type, gender and label of all standards in the kit together with the circuit model parameters 
(offsets, load) or S-parameter tables (.snp file) that are necessary to determine its magnitude and 
phase response.  
To export cal kit data, the analyzer uses a specific binary file format *.calkit.  
Three different import file formats are supported: ZVL-specific binary cal kit files (*.calkit), ZVR-
specific binary cal kit files (*.ck), cal kit files in Agilent-specific ASCII formats (*.csv, *.prn; see 
notes below). 
R&S ZVL 
GUI Reference 
Channel Menu 
Operating Manual 1303.6580.32-06 
243
By default cal kit files are stored in the C:\R_S\Instr\user\Nwa\Calibration\Kits directory.  
To import a ZVL-specific or ZVR-specific cal kit file (*.ck or *.calkit) you can also use the Windows 
Explorer and simply double-click the file or drag and drop the file into the NWA application. The imported 
cal kit file is automatically set active.  
Importing older ZVR cal kit files 
On loading some older ZVR-specific *.ck files, e.g. the ZV-Z23 cal kit file, the ZVL generates the 
message File does not comply with instrument calibration kit file format. The files must be 
converted using a ZVR network analyzer equipped with a firmware version V3.52 or higer. Proceed 
as follows: 
On the ZVR, press CAL – CAL KITS – MODIFY KITS – INSTALL NEW KIT to import the *.ck 
file.  
Press CREATE INST FILE in the same submenu to export the *.ck file in a ZVL-compatible 
format.  
Import the converted file into the ZVL.  
*.csv cal kit files: VNA Cal Kit Manager 2.1 
The VNA Cal Kit Manager is a free, Windows-based software tool intended to import, edit, and 
export *csv cal kit files. The software is available for download at 
http://www.vnahelp.com/products.html. The decimal separator used by the VNA Cal Kit Manager 
V2.1 depends on the language version of the Windows operating system: Cal kit files generated on 
an English operating system contain dots, the ones generated on a German system contain 
commas. 
The network analyzer expects the dot as a separator and displays an error message when a *.csv 
file with commas is loaded. Please install the VNA Cal Kit Manager V2.1 on an appropriate (e.g. 
Englisch) Windows version to avoid trouble.   
*.prn cal kit files: PNA Cal Kit Editor 
The network analyzer can import and process cal kit files created with the PNA Cal Kit Editor. The 
files use the extension *.prn; the data format is identical to the *.csv format.  
The decimal separator used by the PNA Cal Kit Editor depends on the language version of the 
Windows operating system: Cal kit files generated on an English operating system contain dots, 
the ones generated on a German system contain commas. 
The network analyzer expects the dot as a separator and displays an error message when a *.prn 
file with commas is loaded. Please install the PNA Cal Kit Editor on an appropriate (e.g. Englisch) 
Windows version to avoid trouble.   
Remote control:
MMEMory:LOAD:CKIT "file_name"  
MMEMory:STORe:CKIT "kit_name", "file_name" 
[SENSe<Ch>:]CORRection:CKIT:INSTall "<file_name>"
Add or View / Modify Calibration Kit  
Combines different standards to form a new calibration kit, assigning a connector type and further 
attributes (name, labels). This dialog is opened from the Calibration Kits dialog: If the Add Kit... button is 
selected, the dialog title is Add Calibration Kit; if the View / Modify Kit... button is selected, the dialog title is 
View / Modify Calibration Kit. The functionality of both dialog versions is the same. 
If Add Kit is used for a predefined calibration kit, the View / Modify Standard... button can be used to 
R&S ZVL 
GUI Reference 
Channel Menu 
Operating Manual 1303.6580.32-06 
244
open the View / Modify Standard in <kit_name> dialog check the standard parameters. All other 
controls are inactive.  
In its upper part the Add or View / Modify Calibration Kit dialog contains several controls to do the 
following: 
Select a predefined or user-defined Connector Type.  
Insert the (unique) Name of the new kit and assign a Label.
Qualify whether or not the Offset and Load parameters for the standards are defined and 
displayed in Agilent Mode.
Assigning a label to user-defined calibration kits is optional. However, the label is displayed in many 
dialogs and can provide useful information about the kit, e.g. its serial number.  
The table in the center of the dialog lists all standards in the new kit together with either their gender or 
their port assignment (see Restrict Port Assignment) and the label (if defined). Clicking a standard opens 
the View / Modify Standard in <kit_name> dialog where it is possible to change the parameters. Four 
buttons can be used to change the entries in the list. 
Copy Standards from... opens a dialog to copy standards from another calibration kit to the new 
kit.  
Add Standard... selects, defines and adds a new standard to the kit. A calibration kit may only 
contain a single standard of each type.  
View / Modify Standard... displays or changes the properties of the selected standard. 
Delete Standard removes the selected standard from the list and from the calibration kit.  
Use Copy Standards from... to avoid the (re-)definition of their properties. Standards are copied 
together with their parameters and their label, which you can modify once they belong to the new kit.  
Copy Standards from 
Copies standards together with their parameters from a calibration kit in use to a new calibration kit. This 
dialog is opened from the View Calibration Kit dialog (Copy Standards from... button).